Chapter 8
8-8 V60 Ventilator Service Manual 1049766 Rev A
Performance Verification
8.3 View and Record
Ventilator Information
1. If the ventilator is not already in diagnostic mode, press and hold the
nav-ring Enter button, and press the On/Shutdown button on the user
interface. Within 5 seconds of power up, release and press the nav-
ring Enter button again to enter the Diagnostics menu.
2. Touch Service to display the Ventilator Information screen (Figure 8-
1), and record the following information on the Electrical Safety
(section 8.8, Electrical Safety Data Form) or Performance Verification
Data Form (section 8.9, Performance Verification Data Form):
Figure 8-1: Ventilator Information Screen
3. Touch Pneumatics and record the displayed barometric pressure.
4. Measure and record the ambient temperature and relative humidity.
5. Complete a diagnostic report (DRPTA) download (Chapter 6).
6. Turn the ventilator OFF.
Software options
Ventilator serial number
Software version
Total power-on hours
Hours since last PM
Battery lot number
Battery manufacture date