Chapter 7
7-36 V60 Ventilator Service Manual 1049766 Rev A
Component Removal/Installation
7.33 LCD Follow these steps to remove the LCD (Figure 7-34). Reverse to install.
NOTE: When installing a new LCD, remove the protective film before reinstalling
the LCD tray to the front bezel, and avoid touching the LCD screen.
1. Open and separate the UI bezel (section 7.28).
2. Remove the front bezel (section 7.32).
3. Remove the screws that attach the LCD to the LCD tray.
4. With the LCD face up, carefully lift the LCD slightly to disconnect the
backlight inverter and LCD cable.
5. Remove the LCD.
NOTE: When installing the LCD cable to the LCD, use a new piece of Kapton
tape.