EXPOSURE VALUES
36 Planmeca Intra X-ray Unit
User’s Manual
13 EXPOSURE VALUES
13.1 Default exposure values
When the unit is switched on, the default exposure values
appear on the displays.
These values can be programmed by the user, see
section 14.1 “Programming the default exposure and
density values” on page 39.
NOTE The exposure values are programmed corresponding
the density value 0 (factory preset value). The expo-
sure time values are automatically scaled according
to the density value.
NOTE These values are for speed F films (Kodak Insight).
For the speed E films (Kodak Ektaspeed) select 1 step
longer and for the speed D films 4 steps longer expo-
sure time.
NOTE The following exposure values need 8 mA.
NOTE The preprogrammed default exposure values are for
the 20 cm (8”) cone. When using the 30 cm long cone,
use the exposure values given in the table below.
NOTE When using the digital sensor change the exposure
values by pressing the MODE key.
PATIENT kV time
Adult 63 0.080
Child 60 0.064
PATIENT kV time
Adult 63 0.160
Child 60 0.120
PATIENT kV time
Adult 63 0.050
Child 60 0.040
PATIENT kV time
Adult 63 0.160
Child 60 0.125

Other manuals for Planmeca intra

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca intra and is the answer not in the manual?

Planmeca intra Specifications

General IconGeneral
ManufacturerPlanmeca
ModelPlanmeca Intra
Frequency50/60 Hz
Tube Voltage60-70 kV
Tube Current7 mA
TypeDental X-ray Unit
Voltage100-240 V
WeightVaries depending on configuration; refer to specific model documentation

Related product manuals