BEFORE EXPOSURE
8 Planmeca ProSensor
User’s Manual
7.3 Selecting exposure values
Select the digital imaging mode of the unit or adjust the
exposure time according to the table.
NOTE In the digital imaging mode the highest time value
that can be selected is 0.80 seconds.
Table 3. Exposure values for Planmeca ProSensor sensors with 20 cm (8”) cones
I INCISORS
P PREMOLARS AND CANINES
M MOLARS
TIME
0.01s
0.02s
0.03s
0.04s
0.05s
0.06s
0.08s
0.10s
0.12s
0.16s
0.20s
0.25s
0.32s
0.40s
0.50s
0.64s
0.80s
70 kV/
child
I P M maxilla
I P M mandible
66 kV/
child
I P M maxilla
I P M mandible
63 kV/
child
I P M maxilla
I P M mandible
60 kV/
child
I P M maxilla
I P M mandible
57 kV/
child
I P M maxilla
I P M mandible
55 kV/
child
I P M maxilla
I P M mandible
52 kV/
child
I P M maxilla
I P M mandible
50 kV/
child
I P M maxilla
I P M mandible
70 kV/
adult
I P M maxilla
I P M mandible
66 kV/
adult
I P M maxilla
I P M mandible
63 kV/
adult
I P M maxilla
I P M mandible
60 kV/
adult
I P M maxilla
I P M mandible
57 kV/
adult
I P M maxilla
I P M mandible
55 kV/
adult
I P M maxilla
I P M mandible
52 kV/
adult
I P M maxilla
I P M mandible
50 kV/
adult
I P M maxilla
I P M mandible