EasyManuals Logo

Planmeca ProSensor User Manual

Planmeca ProSensor
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
Planmeca ProSensor 21
TECHNICAL SPECIFICATIONS
User’s Manual
A.2 Preprogrammed settings values
For programming these values see the Planmeca ProX
User’s manual, section 16.3 “Programming the
preprogrammed settings”.
NOTE Two sets of exposure values (time/kV/mA) have been
programmed for each exposure region: one for adult
mode and one for child mode.
INCISORS
PREMOLARS AND
CANINES
MOLARS
kV mA time kV mA time kV mA time
Adult
Maxilla 60 7 0.100 63 6 0.1 63 6 0.125
Mandible 60 7 0.08 63 6 0.08 63 6 0.100
Child
Maxilla 60 7 0.063 60 7 0.08 60 7 0.100
Mandible 60 7 0.05 60 7 0.064 60 7 0.080
OCCLUSAL
EXPOSURE
ENDODONTIC BITE-WING
kV mA time kV mA time kV mA time
Adult
Maxilla 70 6 0.100
60
7
0.100
60
7
0.100
Mandible 70 6 0.100
Child
Maxilla 66 6 0.080
60
7
0.080
60
7
0.080
Mandible 66 6 0.080

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProSensor and is the answer not in the manual?

Planmeca ProSensor Specifications

General IconGeneral
BrandPlanmeca
ModelProSensor
CategoryMedical Equipment
LanguageEnglish

Related product manuals