TECHNICAL SPECIFICATIONS
User’s Manual
A.2 Preprogrammed settings values
For programming these values see the Planmeca ProX
User’s manual, section 16.3 “Programming the
preprogrammed settings”.
NOTE Two sets of exposure values (time/kV/mA) have been
programmed for each exposure region: one for adult
mode and one for child mode.
INCISORS
PREMOLARS AND
CANINES
MOLARS
kV mA time kV mA time kV mA time
Adult
Maxilla 60 7 0.100 63 6 0.1 63 6 0.125
Mandible 60 7 0.08 63 6 0.08 63 6 0.100
Child
Maxilla 60 7 0.063 60 7 0.08 60 7 0.100
Mandible 60 7 0.05 60 7 0.064 60 7 0.080
OCCLUSAL
EXPOSURE
ENDODONTIC BITE-WING
kV mA time kV mA time kV mA time
Adult
Maxilla 70 6 0.100
60
7
0.100
60
7
0.100
Mandible 70 6 0.100
Child
Maxilla 66 6 0.080
60
7
0.080
60
7
0.080
Mandible 66 6 0.080