Planmeca ProSensor 19
TECHNICAL SPECIFICATIONS
User’s Manual
Ethernet ControlBox
Dimensions 112 x 46 x 24 mm (4.41 x 1.81 x 0.94 in.)
Power supply 48 V DC 65 mA
Cables
ControlBox to PoE RJ45 10m OR 15m
PoE to LAN RJ45 10m OR 15m
Power supply Phihong Single Port Injector
Type: PSA16U-480 (POE)
Input voltage 100-240 VAC (50-60 Hz)
Output voltage 48VDC
Max. output current 0.35 A
Insulation voltage
Primary-secondary 3000VDC
USB ControlBox
Dimensions 112 x 46 x 24 mm (4.41 x 1.81 x 0.94 in.)
Cables fixed USB 2.0 power supply cable 2 or 5m
(6.6 or 16.4 ft)
Power supply Input power 2 W
Operating environment
Planmeca ProSensor is for indoor use only. The equipment is installed on the wall or on/under the
table. The user moves the sensor into the operation position by hand.
The room and operation must comply with the x-ray safety shielding requirements according to
radiation safety regulation in the country.
The system is used by dental care professionals.
Prior to installation of the system check that the local conditions are compatible with the appliance
design.
The temperature of the operating environment should be between + 15°C and + 40°C.
The relative humidity of the operating environment should not exceed 60%.
Atmospheric pressure range should be between 700 hPa - 1060 hPa.
Transportation and storage environment
Transportation and storage temperature -10°C - +60°C.
The relative humidity during transportation and storage should not exceed 95%.