Rev. 03 / Jan. 2019
Plasmapp Co., Ltd.
STERLINK
®
FPS-15s Plus User Manual
32
5.6. Cycle in progress and completed
In the SR
TM
process, the drying, heating and measuring the residual water are included in the
load condition.
Figure 5.4 GUIs for the drying and sterilization process
Number
Name
Description
14
SR
TM
status
The status image will be blinking during the drying process
15
Sterilization status
The status image will be blinking during the sterilization process.
16
Sterilization time
The remaining sterilization process time is displayed.
When the sterilization process is successfully completed, the smart complete (SC
TM
) process is
performed in order to ensure that there is no residual sterilant left on the sterilized instruments.
After the relatively short SC
TM
process, the sterilization cycle is finished with summary of the
cycle. Press the confirm switch to return to ready state.