9.4. Sterilant cassette errors
The STERLINK
®
FPS-15s Plus sterilization system is designed to initiate automatically the
sterilization cycle according to the barcode information provided by the sterilant cassettes. The
predictable error source is presented with the intuitive figures, and the user should identify the
source and confirm. If the problems persist, contact local Plasmapp customer service
representative.
Figure 9.4 GUIs for the errors related with the sterilant cassettes
The user should confirm the reason of the error and press Confirm
button.
Probable cause: The used cassette is tried to be reused.
Suggestion: Try again with a new cassette.
Probable cause: The expired cassette is tried to be used.
Suggestion: Try again with a new cassette or check time setting.
Probable cause: The barcode of the cassette is invalid.
Suggestion: Try again with a new cassette.
Probable cause: The cassette is not detected.
Suggestion: Clean the barcode window or try again.