5 Operation ■
39
• Printing the application settings
– Keep «PRINT» pressed until "PRINT APPLICATIONS" is displayed.
– Release «PRINT».
– The application settings and all the settings for the methods
currently saved are printed.
5.5.3 The change key
• Changing units
– If drying has already been performed or is under way, the unit of
measurement displayed can be changed using the change key
«
F». Once a drying procedure is ended, the drying result can be
displayed in all the available units until the start of a new drying
sequence using the change key «
F».
– Release «
F» once the unit to which you wish to change is
displayed.
5.5.4 The start/stop key
• Starting drying manually
– Ensure that there is no drying taking place, i.e. that the instrument
is in weighing mode.
– Press «START/STOP» briefly.
– The measurement is started.
• Stopping drying manually
– Drying is under way.
– Press «START/STOP» briefly.
– The measurement is stopped.
Applications :
--------------
Printout of the application settings and
methods.
Dryer setup :
Heat mode : Standard
Time Boost : 3.0 Min
Temperature : 105 C
Timer stop : off
Time : 10.0 Min
Auto stop : Digit/Time
Free : 2/20 D/s
Autostart : off
Standby Temp : off
Temperature : 40 C
Unit : 100-0%
Current drying parameter settings
Targe weight :
Weight check : off
Nominal : 5.000 g
Upper limit : 6.000 g
Lower limit : 4.000 g
Current starting weight settings