6 Determining moisture levels ■
47
6.3.2 Temperature
This function key is used to start the drying temperature input.
The drying temperature can be entered in 1°C increments from 30°C
- 170°C.
The temperature setting for drying with the moisture analyser can be
selected lower than than for drying with the oven drying method.
6.3.3 Timer stop
This function key is used to define the drying time.
If the timer stop is switched on, the measurement procedure is ended
once the time set has elapsed.
The time can be set in 0.1 min increments from 0.1 - 240.0 min.
6.3.4 Switch-off criteria
This function key is used to start the definition of the switch-off
criteria for the measurement.
A freely definable stop mode is available according to the "digit per
time" principle. There is also the automatic stop mode "ADAPT-
• DRYING TEMPERATURE
TEMPERATURE 105°C Temperature input
•DRYING TIME
STOP TIME 10.0 MIN Only if the timer stop is switched on
TIMER STOP ON/OFF