8 Service ■
61
8.3.3 Temperature test
The temperature test is used to verify the temperature measuring
system of the moisture analyser to a known reference temperature
measuring instrument.
The temperature sensor must be inserted in the sample chamber.
Close the cover and follow the procedure:
Temp. Reference ID : ...
Identifier of the temperature adjustment
set
Temperature 100 C : 101 C
Temperature 160 C : 161 C
Values of the temperature adjustment
Temperature Calibration o.k.
Temperature 101 C : 101 C
Temperature 161 C : 161 C
Instrument has been correctly
temperature-calibrated
Operator :
Operator ID, if activated under Set Print
Format (see chapter 5.3.3 "Configuring
the report printout").
Display Key Step
+0,000 9
The balance is in Weighing
mode.
+0,000 9
«T»
Press until "TEST TEMPERA-
TURE" apears.
TEST TEMPERATURE
- - - - - - -
«A»
Set the lower test value.
T100 t160 start
- - - - - - -
«B»
Set the higher test value.
T100 t160 start
Note: The difference between the two temperature-values must be
bigger than 40°C.
- - - - - - -
«START»
Start the test
T100 t160 start