6 Determining moisture levels
48
STOP".
End-point condition Digit per time
The drying procedure ends as soon as the reduction in weight, during
the time set, is smaller than the number of digits set. The reduction in
weight must have been greater than the end-point condition at some
point.
In the freely definable mode, the user can set from 1 - 99 digits in
1 digit increments and from 10 - 90 seconds in 10 second increments.
One digit is the smallest change in measurement that can be displayed
by the moisture analyser.
In the XM 50: 1 digit = 1 mg
ADAPTSTOP
Is a fully automatic stop mode which determines the switch-off time
on the basis of the drying progress.
•STOPMODE
DIGIT/TIME 2/20 Only if Digit/Time is selected
AUTO STOP OFF
DIGIT/TIME
ADAPTSTOP
t [s]
m[Digit]
20 s 60 s
2 Digit (2 mg)
AUTO STOP 2 / 20
AUTO STOP 2 / 60
2 Digit (2 mg)