EasyManuals Logo

Prodigy Preferred User Manual

Prodigy Preferred
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
19
Prodigy Diabetes Care, LLC 1.800.243.2636
Prodigy Meter Owner’s Manual
STEP 3: Apply Control Solution
Shake the control solution bottle
well then remove cap.
Squeeze the bottle and discard the
first drop then wipe off the dispenser
tip with clean tissue paper or cotton.
Squeeze the bottle again to get a
second drop onto a clean,
non-absorbent surface or on
your fingertip. See Figure 16.
Apply the drop to the opening of the
absorbent channel of the test strip (where it meets the narrow
channel) until the confirmation window is filled. See Figure 17
The meter will beep and begin to count down.
See Figures 17 and 18.
Caution: To avoid contaminating the control solution with
the content of the test strip, DO NOT DIRECTLY APPLY
CONTROL SOLUTION ONTO THE TEST STRIP.
Figure 16
P R O D I G Y
Preferred
TM
Figure 17 Figure 18

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Prodigy Preferred and is the answer not in the manual?

Prodigy Preferred Specifications

General IconGeneral
BrandProdigy
ModelPreferred
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals