STEP 4 : Check if the Test Result is in Range.
After seven (7) seconds,
the control solution test
result appears on the
display. Compare the
test result with the range
printed on the test strip
vial. The result should
fall within the printed
range. See Figure 19.
Out of Range Results
If test results fall outside
the range printed on the
test strip vial, check
“Trouble-Shooting Guide”
located in the “System
Troubleshooting” Section
and repeat the test.
If you continue to get out-of-range results, it means that the system
or the control solution may not be working properly. DO NOT use the
system to test your blood glucose level. If you are unable to resolve
the problem, contact Technical Support at 1.800.243.2636.