Page 63 of 68
Proton Products InteliSENS PD30 Instruction Manual (Issue 1a)
DW
Bit
Comments
Unit
Range/Remark
Default value
9
10
11
12
13
14
15
Response speed
0=Instant;1=Average
d
0
DW64
Full scale for analogue O/P
1
1=0.01um/0.01mil
0~20000000
1000000
DW65
DW66
Full scale for analogue O/P
2
1=0.01um/0.01mil
0~20000000
1000000
DW67
DW68
Full scale for analogue O/P
3
1=0.01um/0.01mil
0~20000000
1000000
DW69
DW70
SPC switch
0=OFF;1=ON
0
DW71
Statistics time
1=1s{1=1m{ft}}
1~65535
10
DW72
Statistics Mode
0=Time;1=Length;2=
One Reel
0
DW73
FFT sampling rate
0=OFF;1=1Hz;2=3Hz;
3=10Hz;4=30Hz;
5=100Hz;6=300Hz;7=
1KHz
0
DW74
Hold time for external
alarm
1=1ms
1~5000
300
DW75
DW76
DW77
DW78
DW79
DW80
Profibus ID
1=1
0~125
4
DW81
DeviceNet address
0~63
3
DW82
DeviceNet baud rate
0=125K;1=250K;2=50
0K
0
DW83
CAN BUS address
1=1
0~255
17
DW84
CAN BUS baud rate
0=250;1=500;2=1000;
other=500
0
DW85
RS232 baud rate
0=4800;1=9600;2=19
200;3=38400;4=1152
00
4
DW86
RS232 protocol
0=Modbus;1=Proton;
2=SLP;3=Print
0
DW87
RS422/RS485 protocol
0=Modbus;1=Proton;
2=SLP
0

Table of Contents