Page | 18
Actinic pulse setting
Intensity of Actinic pulse should correspond with cultivation light intensity or should be set according to application.
Should Overflow be observed on the display during the measurement, lower the intensity of the used pulses.
In case of Low value on display during measurement, increase the intensity of the used pulses.
5.2 MEASUREMENT
No device calibration is needed before chlorophyll fluorescence measurements are made. Results of fluorescence measurement depend
on device settings and the samples.
How to perform Chlorophyll Fluorescence measurement with FluorPen:
Prepare dark adapted sample first (prior to F0, QY, NPQ, LC measurements) by placing the sample for at least 10-15 min in the dark.
Alternatively, dark adaptation can be easily achieved by placing the detached leaf clips in closed position on the leaf ahead of the
measurements. Only the FP110/D or PAR-FP110/D is designed for use with the detachable leaf clips. The duration of dark-adaptation
period depends on plant species and growth conditions.
For light adapted measurements no dark adaptation of the sample is required.
Turn ON the device by holding the SET button for 1 sec.
Place the dark-adapted leaf in the leaf-clip (FP110/S or PAR-FP110/S) or in case of detachable leaf clips place the leaf clip on the
optical probe of the FluorPen and slide open the screen of the leaf clip to expose the leaf to the optical probe.
Select Measure > from the menu and select required parameter for example QY (press SET as Enter button when making
selections).
Press SET to start the measurements.
When OJIP, LC or NPQ are being measured the display on the device shows the progress of the measurement as percentage.
When Ft or QY are measured the values appear on the device display. The result of OJIP, NPQ or LC protocol are not visible on the
display of the device and need to be download to PC computer (via USB cable or BT connection, see instructions in Chapter 7 and 8)
using FluorPen Software (downloaded to PC earlier).
All measured data are stored in the device memory and can be downloaded to PC computer after completion of the experiment.
5.3 PROTOCOLS EXPLANATION
5.3.1 OJIP PROTOCOL
The FluorPen device offers the protocol to capture rapid fluorescence transient OJIP, which occurs during exposure of photosynthetic
organisms to high irradiance. The FluorPen software enables data downloading to a PC and subsequent OJIP curve and calculated data
visualization.

Other manuals for PSI FluorPen FP 110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PSI FluorPen FP 110 and is the answer not in the manual?

PSI FluorPen FP 110 Specifications

General IconGeneral
BrandPSI
ModelFluorPen FP 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals