Supervisor ID Error
The supervisor ID
should have between
4-12 characters.
Please check Supervisor
ID and re-enter.
###.###.061.012
Temperature Warning
Temperature is out of
range.
Temperature range
allowed is 15°C to 30°C/
59°F to 86°F.
###.###.061.018
Unknown Test Type Error
The required Test Type
is not installed.
Refer to page 26 to see
instructions for installing
Test Type files.
###.###.061.031
User Type Cannot Be
Changed
This is the last
supervisor. Its type
cannot be changed.
Create a replacement
Supervisor or set a new
default Supervisor
before changing User
status to Operator.
###.###.061.054
Wrong Value Measured
Stray light or wrong
calibration.
Restart the Sofia, and if
issue continues, contact
Technical Support.
###.###.071.021