4
© 2023 Radiacode Ltd.
Introduction
13. Swing button "down", when the display image is turned upside down.
14. Graphical representation of dose rate.
15. Numerical representation of dose rate.
16. Source of audible indication.
17. Units of numerical representation of dose rate.
18. Light indication:
- charge (blue color);
- gamma-quantum registration (green in normal, red when the alarm threshold is exceeded).
19. Random error of dose rate estimation.
1.3 Technical specifications
General technical characteristics
Energy range of registered photon radiation (X-ray and
gamma)
Dose rate assessment:
§
range (by Cs-137)
§
error
§
time (display update period)
§
time for a level of random error ±15% with natural
radiation background 0.08 µSv/h
0.1…1000 uSv/h
±15%
0.5 s
34 s
Accumulated dose assessment:
§
range
Non-volatile memory:
§
registration volume of Standalone mode
Continuous operation time (for fully charged battery)
Built-in battery:
§
type and capacity
§
charging voltage, typical
§
charging current, max
§
full charge time, typical
Li-Pol 3.7 V, 1000 mAh
5 V
500 mA
2 h
Operating temperature range
Wire connection:
§
type
§
cable
§
current consumption, max
USB 2.0 and higher
type C
500 mA
BLE (Bluetooth 4.0 and higher)
Monochrome graphic LCD, 128x48
pixels, 34x13 mm, FSTN,
Transflective, Positive
Dust and moisture protection, code
Requirements for