- 90 - 
 
- command understood but not accessible at this moment 
- command understood and carried out 
Set min checkweighing threshold 
Format:   DH_XXXXX CR LF, where: _ - space, XXXXX – mass format 
- command not recognised (mass format incorrect) 
Set max checkweighing threshold 
Format:   UH_XXXXX CR LF, where: _ - space, XXXXX - mass format 
- command not recognised (mass format incorrect) 
Give value of min checkweighing threshold 
Response: DH_MASA CR LF - command carried out 
1  2  3  4-12  13  14  15  16  17  18  19 
D  H  space  mass  space  unit  space  CR  LF 
-   9 characters with right justification 
-   3 characters with left justification 
Give value of max checkweighing threshold 
Response: UH_MASA CR LF - command carried out 
1  2  3  4-12  13  14  15  16  17  18  19 
U  H  space  mass  space  unit  space  CR  LF 
-   9 characters with right justification 
-   3 characters with left justification 
Set mass value of a single item (only for PARTS COUNTING) 
Format:   SM_XXXXX CR LF, where: _ - space, XXXXX – mass format 
SM_I CR LF 
- command understood but not accessible at this moment (e.g. mode other than 
PARTS COUNTING) 
- command not recognised (mass format incorrect) 
Set target mass value (e.g. for DOSING) 
Format:   TV_XXXXX CR LF, where: _ - space, XXXXX - mass format 
TV_I CR LF 
- command understood but not accessible at this moment (e.g. mode other than 
DOSING) 
- command not recognised (mass format incorrect) 
Set reference mass value (e.g. for PERCENT WEIGHING) 
Format:  
, where: 
space
 
 – mass format 
Response options: 
RM_I CR LF 
- command understood but not accessible at this moment (e.g. mode other than 
PERCENT WEIGHING) 
- command not recognised (mass format incorrect)