Return to weighing:
• Press the
key twice.
15.5.2. Mass of standard inscribed to scale memory
Procedure:
• Enter to <PrcB> function:
• The program goes to the weight display window:
• Using and set standard mass,
where:
- digit selection, - digit setting.
• Confirm the entered value by pressing
,
• You will see the indication equal to 0,000%,
• From this moment display will not indicate the mass of weighed
load but deviation of the load mass placed on the pan referring
mass of standard (in
%).
- 50 -