- 68 -
TO_TARA CR LF
- command executed
Frame format:
1 2 3 4 5-6 7-15 16 17 18 19 20 21
T O space stability space tare space unit CR LF
Tare - 9 characters with decimal point justified to the right
Unit - 3 characters justified to the left
19.4.4. Send the stable result in basic unit
Syntax: S CR LF
Possible answers:
S_A CR LF
S_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
S_I CR LF
- command comprehended but cannot be executed
S_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in basic unit is returned
Frame format:
1 2-3 4 5 6 7-15 16 17 18 19 20 21
S space stability space sign mass space unit CR LF
Example:
S CR LF – computer command
S _ A CR LF - command accepted and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done, mass value
in basic unit is returned.
19.4.5. Send the result immediately in basic unit
Syntax: SI CR LF
Possible answers: