Put a weight of the displayed mass value on the pan and press .
The calibration process will start which is signalled by the message:
After completion of the process of calibration the following screen
will appear
Take off the weight , then the following sequence of screens
will appear
Calibration process can be terminated anytime by pressing
which is signalled by the following message on the display:
Return to weighing with saving changes that have been made.
Caution:
If the calibration process (span adjustment) lasts longer than 15 the
<Err8> message will be displayed and short audible signal will be
emitted. Press
to perform calibration again with more stable
ambient conditions!