Wuhan Raycus Fiber Laser Technologies Co., Ltd
User Guide
RFL-C2000S-HP /RFL-C3000S-HP /RFL-C4000S-HP/RFL-C6000S-HP
26
EMON
Start Emission
Turn on the laser (or enable it)
Send“EMON\r”
Return
“EMON\r”
MPWROFF
Main Power OFF
Turn off the main power
Send“MPWROFF\r”
Return
“MPWROFF\r”
MPWRON
Main Power ON
Turn on the main power
Send“MPWRON\r”
Return
“MPWRON\r”
SPW
Set Pulse Width
Send“SPW 100\r”
Return
“SPW
100\r”
Setpulsewidth100ms
Other return value
“ERR
input Err\r”
input pulse width<0.0001
“ERROut of Range\r”
Exceeds the maximum pulse
width range
“ERRDuty Cycle too High\r”The
set of duty cycle too high
“ERRDuty Cycle too Low\r”Set
pulse width <0.02
“SPW
100,Duty=100%\r”
Working in continuous mode
SPRR
Set Pulse Repetition Rate
Send
“SPRR 1000\r”
Return“SPRR1000\r”
Other return value
“ERR
input Err\r”
Input frequency <0
“ERR Out of Range\r”
Exceeds the maximum pulse
width range
“ERRDuty Cycle too High\r”The
set of duty cycle too high
“ERRDuty Cycle too Low\r”Set
pulse width <0.02
“SPW
100,Duty=100%\r”
Working in continuous mode
SDC
Set Diode Current
Set the operating current of the optical
module (unit:%).The setting value must
be lower than 100% and higher than the
minimum current setting value, which
can be set to 0. If the set value is bigger
than 100, the default is input 100.
Send
“SDC 100\r”
Return“SDC100\r”
Otherreturn value
“ERR
Input Err\r”
input value <0
“Laser is worked in AD Mode\r”
Working in external AD mode,
the command is invalid
RCS
Read Current Setpoint
Send
“RCS\r”
Return“RCS56.7\r”
(Present current setting value is
56.7%)