Preliminary
5 Quick Start Guide
5.1 Connection
The FIRST cable you have to connect and lead to earth ground, is always the Safety GND!
Afterwards connect the HV-GND, the HV-cable and the measurement cable to the test object.
Finally, connect the Safety-Switch to the Instrument.
5.2 Instrument operation
1. Power on the instrument
1
2. Select voltage and frequency
1 The battery symbol on the top of the screen is only available in the CAPO 2.5
90315-1.2 Instruction Manual CAPO
Page 21 of 117
WARNING !
WARNING !
Hazardous voltage may result in severe injury or death!