9.4 GSTg
The GSTg configuration allows unwanted currents, to bypass the measuring circuit and enables
smaller sections of insulation to be tested individually. Only the ground currents are measured,
using a GST-Guard configuration. Current flowing to terminals with the guard connection is not
measured. This configuration is illustrated bellow:
90315-1.2 Instruction Manual CAPO
Page 78 of 117
GSTg A+B