13900-103 Rev. A Page 43
Appendix C - Software Revision List
Revision 1.00 to 1.01 (12/99)
· Correction of VVL camera time out during initial
synchronization.
· Position correction of XYZ assembly before park routine
is invoked.
· Y2k format change to display the correct year on the
print-out.
· Year format correction for the years 2000 through 2009 for
printer output.
Revision 1.01 to 1.02 (2/00)
· Correction of standard deviation calculation.
· Protocol change to add histogram requirements.
· Revision of the calibration curve (linear to quadratic
curve).
· Year format correction for the years 2000 through 2009 for
calibration date.
Revision 1.02 to 1.10 (4/00)
· Added a printer character set for service software.
· Add a function to clear the shot count and clear history
data with service software.
· Added a function to set the digital pot in the applanation
gain circuit.
· Reset printer off-line at boot-up.
Revision 1.10 to 1.12 (8/00)
· Added software delays to the VVL Cameras.
· Decreased XYZ positioning system speed and overshoot.
· Added button push to test XYZ assembly during assembly.
· Added button push to test printer assembly during
assembly.
· Changed logo/name in software to Reichert on most
units.
· Added option to display kPa (mmHg was the only option).
Revision 1.12 to 1.13 (8/00)
· Added RS232 port command to transfer tonometer board
EPROM data to a replacement ckt. Board.
· Added RS232 port command to test L/R detector (beeper).
· Removed 1.12 version delay in VVL cameras.
Revision 1.13 to 1.14 (9/00)
· Added 5 button push for burn-in function.
· Added year correction (date) to service software.
· Service software changes (cal default values, cal type
printout).
Revision 1.14 to 1.15 (12/00)
· Corrected intermittent operation of paper feed after boot-
up.
Revision 1.15 to 1.16 (5/01)
· Correct lock-up condition of motors.
· Added button push for positioning when a travel lock
when a touch or proximity error occurs.
· Update service address for German service shop.
· Added moving to travel lock position when error code
occurs at boot-up.
Revision 1.16 to 1.17 (2/02)
· Add changes to allow testing with the TCT fixture.
· Add PTEST routine to test the XYZ position sensors and
motors with test software.
· Add DISPLAYTEST routine to test the LCD screen during
assembly.
· Correct German language screens.
· Remove LOADDEF command from the service software
which entered default EPROM data.
· Add a software reset to the burn-in sequence to check
for errors.
· Add button sequence for power up during travel lock
error and bad LCD screen conditions.
· Add a non-clearing shot counter.
· Revise the L/R beeper test.
· Add OEM NCT-800 product into software.
· Modify homing timeout period when XYZ cannot go
home.

Other manuals for Reichert AT550

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Reichert AT550 and is the answer not in the manual?

Reichert AT550 Specifications

General IconGeneral
BrandReichert
ModelAT550
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals