- 17 -
Error Possible Cause Corrective Action
Extractor starts,
although a connected
device is not being
used.
Start-up threshold of one suction
hoses is too low.
Check, which suction hose is open, the relevant letter
will be indicated in the display.
Adjust the start-up threshold of the suction hose
(Chapter 5.3.4).
Extractor fails to stop
when a connected
device is switched off
The dust generating device is not
plugged in to the coupler socket on
the extractor and the suction hose
button is switched on.
Start-up threshold is too low
Shut-off delay is too long.
Switch the suction hose button on the extractor off,
and connect the dust generating device to the coupler
socket on the extractor (Chapter 4.7).
Adjust start-up threshold (Chapter 5.3.4)
Set shorter shut-off delay (Chapter 5.3.5).
No satisfactory setting
for the automatic
start-up feature can be
adjusted.
Some older hand pieces do not
JHQHUDWHDVXI¿FLHQWO\VWURQJVLJQDO
for the automatic start-up feature.
Switch to higher revs per minute (e.g. 10,000 rpm.)
3DQHOVFRYHULQJWKH¿QH
SDUWLFOH¿OWHUDQGWKH
GXVWGUDZHUDUHGLI¿FXOW
to open.
Lock bolts are dirty. Clean lock bolts and lubricate them slightly if
necessary.
The extraction force is
not readjusted when a
second suction hose is
switched on or off
The automatic extraction force
readjustment is deactivated.
Activate the automatic extraction force readjustment
(See chapter 5.2.2).
The display indicates
“OFF” above the button
(41)
The automatic extraction force
readjustment is deactivated.
Activate the automatic extraction force readjustment if
required (See chapter 5.2.2).
There is no audible
signal.
The audible signal is deactivated. Activate audible signal (Chapter 5.3.2)
Pinch valves close /
open slowly
7KH¿OWHULQWKHFRPSUHVVHGDLU
connection is blocked
&KDQJHWKH¿OWHULQWKHFRPSUHVVHGDLUFRQQHFWLRQ
(see chapter 6.2.4).
7.1 Error codes
,IWKHIROORZLQJHUURUFRGHDSSHDUVGXULQJRSHUDWLRQRUVHOIGLDJQRVLVSOHDVHFRQWLQXHDVIROORZV
Switch device off
Deal with the fault as described in the table
Switch the device back on
For all other error codes, switch the unit off and then back on again.
,IWKHHUURUFRGHLVUHSHDWHGO\LQGLFDWHG
Make a note of the error code.
Contact the repair service
Error Code Cause / Meaning Corrective Action
Cod01 7KHVHWGXVWEDJ¿OOOHYHOKDVEHHQ
reached.
Change dust bag (Chapter 6.1).
Err01 6XFWLRQÀRZLVEORFNHG
,QVXI¿FLHQWFRPSUHVVHGDLUSUHVVXUH
Check the suction hoses for blockages, examine
WKHGXVWEDJ¿OOOHYHODQGFKDQJHLIQHFHVVDU\
6HHFKDSWHUH[DPLQHWKH¿QHSDUWLFOH¿OWHUIRU
contamination, and change if necessary (see chapter
6.2.1).
If the device is supplied with compressed air from
a main laboratory switch together with the air
FRPSUHVVRUWKHQDW¿UVWWKHFRPSUHVVHGDLUZLOOEH
LQVXI¿FLHQWDQGWKHVHOIGLDJQRVLVZLOOLQGLFDWHDIDXOW
6ZLWFKWKHGHYLFH21DWWKH2Q2IIVZLWFK¿J
Err04 Overheated electronics Switch the extractor off and allow cooling. If the
extractor is located in a closed cabinet, make sure
the exhausted air is able to escape (See chapter 4.2).
EN