6
Parts checklist
The following services / equipment are required
to facilitate full system functionality:
Tool setting system – fixed or
separate
Ensure that the NC4 system supplied has the
correct separation (for details, see “Access panel
identification markings” on page 7. Contact your
supplier if a different range is required.
Mounting
If mounting brackets are required, ensure they
are correct for the installation.
Air supply
The NC4 requires a clean, dry air supply to ISO
8573-1:Air quality of class 1.7.2. If this is not
achievable a filter/regulator will be required (see
“Parts list – air supply accessories” on page 78).
General
Interface
The NC4 requires the use of the NCi-4 interface.
Software
Renishaw non-contact software is required to
conduct tooling probing / measurement cycles.
Accessories
Equipment such as conduit, fittings, etc. may be
required, dependent upon the installation.
One of the following items of equipment is also
needed for setting up and aligning the NC4
system.
● Digital voltmeter
● NC4 set-up tool