11.2 Getting Started
The KrosFlo System is ready to run after all Mechanical and Electrical Installation steps have been
completed.
To start up the System:
1. Make sure that the Emergency Stop button is pulled to OUT (inactive) position.
2. Rotate the Main Power Disconnect for 90 degrees clockwise.
3. The computer starts and the control software loads automatically.
4. Press the Reset button on Control Panel.
5. Once the computer is booted up, the system is set to automatically logon a special
named account without any user interaction.
6. If a different user is required for logon, wait for the system to completely boot and logon,
and then use the logout function of Secure Desktop to exit the user session to return to a
Windows Login Screen.
7. Once the PC boots up, wait at least 2 to 5 minutes before opening the HMI application
(InTouch WindowViewer). This is to allow the various AVEA (Wonderware) software
components and services to all load up in the background. If InTouch is opened too soon,
the application will not connect to the underlying
AVEVA® software components and not
function correctly.
8. Select the Process icon to navigate to the Process screen.
Note:
Press “Logout” first if the System is still shown as being logged-in by another user.
At startup, the KrosFlo System will display all alarms as being active; they should be
acknowledged, and a reset should be performed to clear any latched faults before
operating the KrosFlo System.
9. To log-in to the System, select Login, enter a pre-configured user account name and
password, and then press OK.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN KrosFlo SYITF-2000 and is the answer not in the manual?

REPLIGEN KrosFlo SYITF-2000 Specifications

General IconGeneral
BrandREPLIGEN
ModelKrosFlo SYITF-2000
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals