Managing Climate Control
Designed to be ideal for most patients, Climate Control Auto enables the automatic delivery of a constant
value of absolute humidity while protecting against rainout.
To allow for increased flexibility, Climate Control can be turned to Manual in either the Patient Setup
(when enabled) or the Clinical Setup menus. Setting Climate Control to Manual disables the automatic
control of humidity and allows the patient to set humidity and temperature levels independently.
However, rainout protection is not provided when Climate Control is set to manual.
Mode
Humidity Temperature
Setting range Default settings Setting range Default settings
Climate Control – Auto Climate Control
Constant
absolute humidity
(depending on
temperature
setting)
Off,
1
60ºF–86ºF
(16ºC–30ºC)
80ºF (27ºC)
Climate Control – Manual
Humidity level Heated tube
Off–6.0
(0.5 increments)
3 Off, 60ºF–86ºF
(16ºC–30ºC)
80ºF (27ºC)
1
When the temperature setting is set to Off, the tube will not heat the air nor will the humidifier heat the water to add
humidity to the air.
368220.indb 19 6/09/2011 9:21:49 AM

Other manuals for ResMed S9 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed S9 Series and is the answer not in the manual?

ResMed S9 Series Specifications

General IconGeneral
Data StorageSD card
Device TypeCPAP Machine
Pressure Range4 to 20 cm H2O
Sound Level24 dB
Power Supply100-240V AC, 50/60Hz
Therapy ModesCPAP, VPAP ST, VPAP Auto
Altitude AdjustmentYes
Operating Temperature+41°F to +95°F (+5°C to +35°C)
Operating Humidity10% to 95% non-condensing
AC Voltage100-240V AC
Humidifieroptional
RampYes