EasyManuals Logo

ResMed S9 Series Clinical Guide

ResMed S9 Series
37 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
5
English
Operating information
The following table describes the operating modes available on the S9 Series.
Mode S9
AutoSet
S9
Elite
S9 Escape
Auto
S9
Escape
CPAP mode
A fixed pressure is delivered with optional
Expiratory Pressure Relief (EPR). EPR can
help ease any discomfort that your patient
experiences when breathing out during
treatment.
ü ü ü ü
AutoSet mode
The AutoSet algorithm automatically adjusts
pressure in response to inspiratory flow
limitation, snore and apnea. AutoSet mode
offers optional EPR. You can set the minimum
and maximum allowable treatment pressures.
ü
APAP mode
APAP mode treats apneas up to 10 cm H
2
O and
does not differentiate between different types
of apneas. The APAP algorithm automatically
adjusts pressure in response to inspiratory flow
limitation, snore and apnea.
ü
EPR technology
Designed to make therapy more comfortable, EPR maintains optimal treatment for the patient during
inhalation and reduces the delivered mask pressure during exhalation. You can select EPR in CPAP,
AutoSet or APAP modes.
EPR checklist
ü You can set EPR to Off, Ramp Only, Full Time.
ü You can select an EPR Level of Off, 1, 2, 3 cm H
2
O or Patient (configurable by the patient).
ü In CPAP mode you can set the EPR Inhale to Med or Fast.
ü When EPR is enabled, the delivered pressure will not drop below a minimum pressure of 4 cmH
2
O,
regardless of the settings.
368220.indb 5 6/09/2011 9:21:26 AM

Other manuals for ResMed S9 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed S9 Series and is the answer not in the manual?

ResMed S9 Series Specifications

General IconGeneral
Data StorageSD card
Device TypeCPAP Machine
Pressure Range4 to 20 cm H2O
Sound Level24 dB
Power Supply100-240V AC, 50/60Hz
Therapy ModesCPAP, VPAP ST, VPAP Auto
Altitude AdjustmentYes
Operating Temperature+41°F to +95°F (+5°C to +35°C)
Operating Humidity10% to 95% non-condensing
AC Voltage100-240V AC
Humidifieroptional
RampYes

Related product manuals