3.3 Electromagnetic Compatibility (EMC)
EMC class according to DIN EN 55011: A
Strong electromagnetic interference fields, such as high-power radio transmitters, can have an
adverse influence on the amplitude control of the AS 200 control. Once the source of the
interference is eliminated, the AS 200 control will return to normal operation by itself.
3.4 Rated Power
~ 460 W (VA)
3.5 Dimensions and Weight
Height without sieve clamping unit: 212 mm
Height with sieve clamping unit: 842 mm
Height with long sieve clamping unit: 951 mm
Width: 417 mm
Width with "comfort" clamping unit: 448 mm
Depth: 384 mm
Weight without sieve stack, without clamping unit: ~ 35 kg
3.6 Required Floor Space
Falling down of the device
Incorrect positioning or insufficient working space
Due to its weight, the device can inflict personal injury if it falls down.
• The device must only be operated on a sufficiently large, strong and
stable workplace.
• All feet of the device must be positioned securely.
Width of the base: 450 mm
Depth of the base: 450 mm
No safety clearances required
Location requirements:
The device must be placed on a vibration-free, plane, stable and free surface to avoid
transmission of vibrations. A level base ensures the uniform distribution of the sample over the
sieve mesh fabric, as well as the stability of the device.
3.7 Receptacle Volume
The maximum receptacle volume (the maximum feed quantity) depends on various factors such
as number and aperture size of the test sieves, maximum grain size and width of distribution of
the sample material.
Examples for the maximum feed quantity according to DIN 66165 for test sieves of 200 mm in
diameter are listed in the following table: