Controlling the Device
30
7.3 Stop Process
The sieving process will stop automatically after the set process time has elapsed. However, the
sieving process can be stopped manually at any time.
Press the button (H1) to stop the sieving process.
By pressing the button, the sieving process stops, the red LED lights up and the green
LED of the button (H2) turns off.
Press the button (H1) a second time to put the device into standby mode.
7.4 Pause Process
The sieving process will stop automatically after the set process time has elapsed. However, the
sieving process can be interrupted manually at any time.
Press the button (H3) to interrupt the sieving process.
The process time is stopped and the green LED of the button (H3) flashes.
Continue the process:
Press the button (H2) to continue with the sieving process.
End the process:
Press the button (H1) to end the sieving process.
7.5 Amplitude
The amplitude display (H4) shows the set amplitude value depending on the setting either in
mm or g (acceleration of gravity). The amplitude value is adjustable between 0.20 mm (1 g) and
3.00 mm (25.5 g). When the device is switched on, the last used amplitude is preset.
Press the "+" or "-" button to set the desired amplitude.
Press and hold the "-" or "+" button to increased or decreased the amplitude in steps of
0.1 mm (1 g), respectively.
The amplitude can also be changed during operation by pressing the "+" or "-" button. An
exceeding or falling below of 3.00 mm or 0.20 mm respectively is not possible.
The amplitude can either be displayed as the height of lift in mm or as a multiple of gravity g
(9.81 m/s
2
), the so called gravitational acceleration.
Simultaneously press the "+" and "-" button to toggle the display between "mm" and "g".
During the sieving process, the amplitude is kept constant within a predetermined tolerance of
0.1 mm.