Wet Sieving
37
8 Wet Sieving
WARNING
W6.0001
Danger to life through electric shock
Wet sieving
An electric shock can cause burns, cardiac arrhythmia, respiratory arrest,
as well as cardiac arrest.
Never operate the device in a water drain basin!
Do not touch the device, if water has entered the interior!
Always operate the device with a mains socket protected by a
residual current circuit breaker (RCCB).
WARNING
W7.0008
Danger to life through electric shock
Ingress of water if the mains plug is not completely plugged in
Water can enter the IEC socket and cause an electric shock if the mains
plug is not completely plugged in.
Only operate the device with the mains plug fully plugged in.
NOTICE
N15.0049
Damage to the sieve mesh fabric
Fluid retention during wet sieving
Fluid retention can lead to overload and therefore to the damage or
destruction of the sieve mesh fabric.
Observe the recommended flow rate.
Always dose the quantity of liquid applied in such a way that no fluid
retention can occur.
Use venting rings, if necessary.
8.1 Installing the Slash Protection
CAUTION Never carry out a wet sieving without mounted splash protection!
The splash protection (SP) is supplied together with the wet sieving lid.
Fig. 15: Installing the splash protection
LS
SP