Packaging, Transport and Installation
15
3.8 Compressed air required
A compressed air connection with at least 7.5 bar is required.
3.9 Electrical Connection
Risk to life caused by an electric shock
Connection to socket without a protective earth conductor
− Connecting the device to sockets without a protective earth conductor can
lead to life-threatening injuries caused by an electric shock.
• Always operate the device using sockets with a protective earth
conductor (PE).
Electrical connection
Failure to observe the values on the type plate
− Electronic and mechanical components may be damaged.
• Connect the device only to a mains supply matching the values on
the type plate.
WARNING When connecting the power cable to the mains supply, use an external fuse that
complies with the regulations applicable to the place of installation.
− Check the type plate for details on the necessary voltage, frequency, and maximum
external current source fuse for the device.
− The listed values must agree with the existing mains supply.
− Only use the supplied power cable to connect the device to the mains supply.
3.10 Transport
Risk of injury due to the device falling down
Lifting the device above head height
− The device can fall causing serious injuries when lifted above head
height.
• Never lift the device above head height!
Damage to components
Transport
− Mechanical or electronic components may be damaged during transport.
• The device must not be knocked, shaken or thrown during transport.