Cleaning, Wear and Maintenance
40
Fig. 32: Position of lubrication points
The drive of the vibratory disc mill has 6 lubrication points (PS) which must be lubricated
regularly after 150 hours of operation.
Use a natural coloured, lithium soap grease (graphite free) such as Shell Gadus S2 V220 2 or
BP Energrease LS-EP 2.
7.2 Maintenance
WARNING
W5.0000
Risk of loss of life due to an electric shock
High voltage due to capacitor discharge
Due to the capacitor discharge on the frequency converter, the device
conducts voltage for up to 3 minutes after the mains plug has been
removed.
You could touch live contacts when the device is open. An electric
shock can cause serious injuries in the form of burns and cardiac
arrhythmia, respiratory failure or cardiac arrest.
Wait 3 minutes after removing the power cable before opening the