Technical Data
21
4.7 Degree of Protection
IP40
4.8 Protective Equipment
The device is fitted with a safety mechanism that prevents the device being started in an
unsafe state.
The device can only be started when the lid is closed.
The lid can only be opened when the device has come to a halt.
The device can only be started if there is sufficient air pressure (at least 7.5 bar).
If the air pressure drops below the minimum level during grinding, the device is automatically
switched off.
4.9 Dimensions and Weight
Height: up to approx. 1420 mm / width: 1150 mm / depth: up to approx. 800 mm
Weight : RS 300 net approx. 315 kg
4.10 Required Floor Space
Height (cover open): 2100 mm / width: 1150 mm /
depth: 800 mm;