Cleaning, Wear and Maintenance
41
Fig. 33: Removing the cover
Fig. 34: Key
Using the key supplied, release the locks (VE) shown in the image.
Remove the cover.
NOTICE
N15.0000
The top and bottom of the springs must be checked every 200 operating hours for wear and
tear or brittleness.
A forklift or similar is needed when replacing the spring in order to lift the top platform.
Fig. 35: Positions of rubber springs
NOTICE
N16.0000
Check the filter element (B) on the filter regulator every 2 weeks. Drain any condensate
present into a separate container and then close the container.
Replace the filter system (B) every 2 years or when the pressure drop reaches 0.1 MPa in
order to prevent damage to the filter element.
VE