Cleaning, Wear and Maintenance
41
Fig. 33: Removing the cover
• Using the key supplied, release the locks (VE) shown in the image.
• Remove the cover.
– The top and bottom of the springs must be checked every 200 operating hours for wear and
tear or brittleness.
– A forklift or similar is needed when replacing the spring in order to lift the top platform.
Fig. 35: Positions of rubber springs
– Check the filter element (B) on the filter regulator every 2 weeks. Drain any condensate
present into a separate container and then close the container.
– Replace the filter system (B) every 2 years or when the pressure drop reaches 0.1 MPa in
order to prevent damage to the filter element.