6 MOUNTING
Caution!
The safety notices contained in Chapter 1 must be complied with!
6.1 Connection of Water Piping
The hydraulic connections are located at the bottom on the left side of the unit The nominal pipe
widths and fitting connections are intended for use of the original hose set and/or the "Active hose
set".
In the case of differing connection conditions, all applicable technical data have to be submitted to
Glen Dimplex Deutschland GmbH in order to obtain from them a written special approval prior to
commissioning. To avoid any damage, the corrosion behaviour of the materials (used in the
equipment to be cooled) with respect to the water must be taken into consideration (see also the
Chapter Commissioning).
Connection of Water Piping
Notice!
Where the piping system is installed at the installation site by the customer, care
must be taken that the pipe system is free from contamination (lines may have to
be flushed prior to connecting the P030 Chiller).