RIGOL Chapter 6 Specifications
User’s Guide for DM3068
General Specifications
Display 256×64 LCD, dual display, Chinese/English menu and online help.
Power Supply AC 100 V – 120 V, 45 Hz – 440 Hz
AC 200 V – 240 V, 45 Hz – 66 Hz
Detect the power frequency automatically at power-on, 400 Hz defaults to 50 Hz
Power Consumption 25 VA Max
Working Environment
Full accuracy for 0℃ to 50℃
Full accuracy to 80% R.H. at 40℃ Non-coagulation
Storage Temperature
-40℃ to 70℃
Operation Altitude Up to 2000m
Safety IEC 61010-1; EN 61010-1; UL 61010-1; CAN/CSA-C22.2 No. 61010-1
Measurement CAT I 1000V/CAT II 300V
Pollution Degree 2
EMC EN 61326-1
Weight About 3.2 kg (without package)
Dimension (height×width×length): 107.0 mm×231.6 mm×290.5 mm
Remote Interface GPIB, 10/100Mbit LAN, USB 2.0 Full Speed Device & Host (support USB storage device), RS-232C
Programming Language SCPI
LXI Compatibility LXI Class C, Version 1.2
Warm-up Time 90 minutes