EasyManua.ls Logo

Rigol DM3068

Rigol DM3068
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Front Panel Operations RIGOL
Users Guide for DM3068
2-3
To Set the Resolution
DM3068 holds reading resolutions of, 4½, 5½ and digits. It automatically
selects a reading resolution according to the current measurement settings. The
greater the resolution is, the higher the measurement accuracy will be; the lower
the resolution is, the faster the measurement will be. Different measurement
function has different resolution.
1. In DCV, DCI and OHM measurements, press the left and right direction keys at
the front panel (
) to set the reading resolution to 5
1
/
2
or 6
1
/
2
digits. The
resolution affects the integration time and vice versa.
Table 2-1 Relationship between reading resolution and integration time
Resolution Integration time Status bar display
5
1
/
2
0.006 PLC
0.02 PLC
0.06 PLC
0.2 PLC
1 PLC
0.006
0.02
0.06
0.2
1
6
1
/
2
2 PLC
10 PLC
100 PLC
2
10
100
2. In ACV and ACI measurements, the resolution is fixed at 6
1
/
2
digits. Press the
left and right direction keys to set the filter type.
Table 2-2 Relationship between the frequency and speed of AC filter
Resolution
AC filter
Speed
Status bar display
6
1
/
2
200 Hz Fast 200 Hz
6
1
/
2
20 Hz Mid 20 Hz
6
1
/
2
3 Hz Slow 3 Hz
3. In FREQ and PERIOD measurements, the resolution is fixed at 6
1
/
2
digits.
Press the left and right direction keys to set the gate time.

Table of Contents

Other manuals for Rigol DM3068

Related product manuals