RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DS1000/DS1000CA Series
2-33
3. :DISPlay:PERSist
Command Format:
:DISPlay:PERSist{ON|OFF}
Explanation:
The command is to set the waveform persistently ON or OFF.
Query Format:
:DISPlay:PERSist?
Returned Format:
Query returns ON or OFF.
Example:
:DISPlay:PERSistON Set the waveform persist ON
:DISPlay:PERSist? Query returns ON
4. :DISPlay:MNUDisplay
Command Format:
:DISPlay:MNUDisplay<time>
Explanation:
The command is to set the display time of the menu. The options are: 1s, 2s, 5s,
10s, 20s or INFinite.
Query Format:
:DISPlay:MNUDisplay?
Returned Format:
Query returns 1s, 2s, 5s, 10s, 20s or INFinite.
Example:
:DISPlay:MNUDisplay10 Set the display time of the menu to 10s
:DISPlay:MNUDisplay? Query returns 10s