Programming Introduction RIGOL
Programming Guide for DS1000CA Series
5. :TRIGger:VIDEO:SENSitivity
Command Format:
:TRIGger:VIDEO:SENSitivity <count>
:TRIGger:VIDEO:SENSitivity?
Function Explanation:
The commands are used to set and query the trigger sensitivity of VIDEO
trigger. The range of <count> is 0.1~1.
Returned Format:
Query returns the set value of <count> in scientific numeric notation in div.
Example:
:TRIG:VIDEO:SENS 0.5 Set the trigger sensitivity as 0.5div
:TRIG:VIDEO:SENS? Query returns 5.00e-01