RIGOL Programming Introduction
Programming Guide for DS1000CA Series
:TRIG:ALT:SLOP:WIND PA Set the border of the vertical window as “PA”
:TRIG:ALT:SLOP:WIND? Query returns “P_WIN_A”
13. :TRIGger:ALTernation:SLOPe:LEVelA
Command Format:
:TRIGger:ALTernation:SLOPe:LEVelA <value>
:TRIGger:ALTernation:SLOPe:LEVelA?
Function Explanation:
The commands are used to set and query the upper trigger level A of Slope
trigger. The range of <value> is -6*Scale~+6*Scale. “Scale” indicates the
current vertical scale, the default unit is V/div.
Returned Format:
Query returns the set value of <value> in scientific numeric notation in V.
Example:
:TRIG:ALT:SLOP:LEVA 2 Set the trigger level A as 2V
:TRIG:ALT:SLOP:LEVA? Query returns 2.000e+00
14. :TRIGger:ALTernation:SLOPe:LEVelB
Command Format:
:TRIGger:ALTernation:SLOPe:LEVelB <value>
:TRIGger:ALTernation:SLOPe:LEVelB?
Function Explanation:
The commands are used to set and query the lower trigger level B of Slope
trigger. The range of <value> is -6*Scale~+6*Scale. “Scale” indicates the
current vertical scale, the default unit is V/div.
Note: The minimum of Level A couldn’t be less than maximum of
Level B.
Returned Format:
Query returns the set value of <value> in scientific numeric notation in V.