EasyManuals Logo

RKI Instruments SP-220 TYPE F User Manual

RKI Instruments SP-220 TYPE F
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
- 16 -
NOTE
When powering on after leaving the detector for more than five minutes with the batteries removed,
such as when powering on for the first time, replacing the batteries, etc., or powering on with the
batteries inserted with incorrect polarities, a clock abnormality (FAIL CLOCK) may be triggered.
When it is reset using the MODE button, the detector moves to the date/time setting screen. See
"6-2. Date/time setting" on page 35 and set the date/time.
Warm-up time (WARM UP) is different depending on the stability status of the sensor.
If the detector is not used for a long period, warm-up time may be longer or a sensor abnormality
(FAIL SENSOR) may be displayed. In such a case, restart the detector.
The detector performs air calibration automatically during start-up. If an environment where the
detector is turned on is different from the atmosphere being measured, perform air calibration again
in the measured atmosphere.
About LCD backlight
Pressing the AIR, MODE or POWER button turns on the LCD backlight. The LCD backlight goes off
after 30 seconds of no operation.
When an alarm is triggered, the LCD backlight lights up automatically.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments SP-220 TYPE F and is the answer not in the manual?

RKI Instruments SP-220 TYPE F Specifications

General IconGeneral
BrandRKI Instruments
ModelSP-220 TYPE F
CategorySecurity Sensors
LanguageEnglish

Related product manuals