9-3. Preparing for a Span Adjustment (Demand Flow
Regulator)
This section describes how to prepare for calibration if you are using a calibration kit that includes:
Demand flow regulator
Calibration tubing
Calibration cylinder
1. Screw the demand flow regulator onto the calibration cylinder.
2. Connect the tubing to the demand flow regulator’s hose barb fitting.
9-4. Automatic Span Adjustment
These instructions describe an automatic calibration of the SP-220. Be sure you have performed a fresh air
adjustment and prepared for calibration as described in the previous sections before continuing.
1
If you have just
performed a fresh air
adjustment and the
instrument is still
running in Calibration
Mode, continue to step
5.
If your instrument is not
currently running in
Calibration Mode,
continue to step 2.
2 With the instrument
running in Detection
Mode, press and hold
the AIR and MODE
buttons for 3 seconds.
3
If the Calibration Mode
password parameter
has been set to on
(factory setting is off),
you will be prompted for
a password.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments SP-220 TYPE F and is the answer not in the manual?

RKI Instruments SP-220 TYPE F Specifications

General IconGeneral
BrandRKI Instruments
ModelSP-220 TYPE F
CategorySecurity Sensors
LanguageEnglish

Related product manuals