2-4. Safety information (for ATEX/IECEx specifications)
Portable gas leak detector SP-220 is a gas detector to continuously detect combustible gases.
The detector draws gases with the internal micro pump.
Use only two TOSHIBA AA alkaline batteries (LR6T) connected in series for power supply.
Do not replace the dry batteries in a hazardous location.
Explosion-proof specifications
Ex ia IIC T4 Ga
Operating temperature range: -20 to +55°C
Electrical specifications
Only TOSHIBA AA alkaline batteries (LR6T) can be used.
Certificate Number
IECEx: IECEx PRE 15.0060
ATEX: Presafe15 ATEX7188
Applied standards
IEC 60079-0:2011 EN60079-0:2012
IEC 60079-11:2011 EN60079-11:2012
Precautions
Do not replace the battery unit in a hazardous location.
Do not replace the dry batteries in a hazardous location.
Do not disassemble/modify the unit.
Use only TOSHIBA AA alkaline batteries (LR6T) for the dry battery unit.
INST. No. 0 0 0 0 0 0 0 0 0 0 0
A B C D E
A: Manufacturing year (0-9)
B: Manufacturing month (1-9, XYZ for Oct.-Dec.)
C: Manufacturing lot
D: Serial number
E: Factory code
Manufacturer
RIKEN KEIKI CO., LTD.
2-7-6 Azusawa, Itabashi-ku, Tokyo, 174-8744 Japan
Web site: http://www.rikenkeiki.co.jp/
NOTE
TOSHIBA AA alkaline batteries (LR6) can be used for TIIS specifications.
II 1 G Ex ia IIC T4 Ga