- 47 -
Use the AIR and MODE
buttons to adjust each
character and press the
POWER button to move
to the next character.
4
The AIR CAL menu item
will be displayed.
5
Use the AIR or MODE
button to scroll to the
AUTO CAL menu item.
6
Press and release the
POWER button. The auto
cal value will be
displayed.
7
To adjust the auto cal
value:
a. Press and hold the
MODE button then
press and hold the
POWER button. The
auto cal value will
begin to flash.
b. Use the AIR and
MODE buttons to
adjust the value.
c. Press the POWER
button to confirm the
change.