- 21 -
NOTE
For switching to the differential measuring mode, see "Change of the measuring mode" in the display
mode on page 31.
CAUTION
Performing an air calibration in the presence of gas will decrease the bar meter.
For excessive temperature/humidity changes, the bar meter may increase.
When the mode is switched from the differential mode to the concentration measuring mode,
check the atmosphere for freshness and perform air calibration again in the measured
atmosphere.
The gas concentration is displayed in six levels (without a unit).
Air calibration is performed automatically and it cannot be performed manually.