22
Control Tests
4
3
Insert the metallic end of
the test strip into the
meter. Place the meter on
a flat surface.
The meter turns on.
Preparing to test
appears.
Apply drop appears.
4
Select the control solution
to test. You will enter the
level later in the test.
5
Remove the bottle cap.
Wipe the tip of the bottle
with a tissue. Squeeze
the bottle until a tiny
drop forms at the tip.
6
Touch the drop to the
yellow edge of the test
strip. Do not put control
solution on top of the test
strip.
Analyzing appears when
there is enough control
solution in the test strip.
7
Control Result and the
control bottle symbol
appear.
Press
or to select
the control level you
tested. If you do not select a level, the
control result is saved without a control
level.
Press
.
8
or
Within range and appear if the control
result is within range.
Out of range and
appear if the control
result is out of range.